California Instruments 148
User Manual Compact i/iX Series
Number Message String Cause Remedy
45 "Current self test error, No. 3/C amplifier in Aux.
phase models only.
output 3 Source has no output
during Self-test. Three
Contact CI service department at
support@calinst.com
216 " RS-232 receiver Communication failure. SB/LAN port settings and cable.
framing error"
Check U
217 " RS-232 receiver parity
error"
ation Communic failure. Check USB/LAN port settings and cable.
218 " RS-232 receiver
overrun error"
Communication failure. Check USB/LAN port settings and cable.
402 "CAL password is
incorrect"
Calibration password does
not equal high voltage
range value.
Re-enter correct password.
403 "CAL not enabled" No password entered for
calibration
Enter correct CAL password.
600 "Systems in mode:list
lengths"
have different list .
Transient lists have
unequal lengths
Check list settings and correct to same
no of data points
601 "Requested voltage and
waveform exceeds peak
voltage capability"
and
e to
exceed peak voltage
capability.
Wave shape selected
RMS voltage combin
Reduce RMS or crest factor of wave
shape.
602 "Requested voltage and
waveform exceeds
transformer volt-second
f he waveform
and – half
rating"
The selected wave shape
exceeds output
transformer capability.
The volt-second product o
(magnitude and time in the +
of wave form).
603 "Command only applies
to RS-232 interface"
nt for
GPIB interface.
Command not releva Do not use command.
604 "Trigger received before
requested number of pretrigger
readings"
t.
Data acquisition pre- r
trigger buffer not filled ye
Hold off trigger or reduce pre-trigge
delay.
605 "Requested RMS current
range"
rrent is t limit
too high for voltage
Max RMS cu
function of voltage range
selected.
Reduce programmed RMS curren
or select low voltage range.
606 "Waveform data not orm name Specify waveform name before sending
defined"
No wavef
specified waveform data.
607 LT:SLEW, and
le"
"VOLT,VO or change waveform type.
FUNC:SHAPe modes
incompatib
Conflict between wave
shape and programmed
slew
Reduce slew
608 "Measurement
overrange"
Measurement data out of
range.
609 "Output buffer overrun" Too much data in output n
buffer.
Check receive mode on applicatio
program. Program is not reading data
sent by AC source.
610
given with present
Check configuration for available options
and features.
"Command cannot be
SYST:CONF setting"
Command conflicts with
available hardware or
firmware option settings.
801 "Output volt fault"
match program value
ltage
s in
e of
L setting.
Output voltage is driven above
programmed voltage by external
- Output voltage does not
when ALC is on.
- Over load
- Voltage kick-back
- No output vo
Load exceeds current limit and unit i
Constant Voltage (CV) mod
operation.
- Reduce load or increase C
California Instruments 149
User Manual Compact i/iX Series
Number Message String Cause Remedy
influence (Load, voltage kickback, etc.)
802 "Current limit fault" limit exceeded. Load exceeds current limit and unit is in
crease CL
Current
Constant Voltage (CV) mode of
operation. Reduce load or in
setting
803 "Temperature fault" Amplifier heat sink temp.
too high.
d
t clearance. Check fan(s) for
Reduce load. Ensure proper airflow an
exhaus
operation.
804 "External sync error" Could not sync to external
sync signal.
External sync signal missing,
disconnected or out of range.
805 "Initial memory lost" Initial settings could not b
recalled at power-up.
e Save power on settings again to
overwrite old content.
806 "Limit memory lost" a
recalled at power-up.
Contact CI service department at
support@calinst.com
Hardw re configuration
settings could not be to obtain
instructions on restoring configuration
data.
807 "System memory lost" Memory corrupted during
 