曝光台 注意防骗
网曝天猫店富美金盛家居专营店坑蒙拐骗欺诈消费者
have a negative error number, status messages have a positive
The table below displays a list of possible error and status mess
and remedy.
error from either the front pan
ges are displayed on
l or the remote control interface wil
CD display. They are al
error ssage queue from wh they can be queried us the SYST:ERR? Query. The error
than can be held in the queue, a
o empty the queue, use the error
ry unt he No Error res ved.
tiple e r messages are g d in succession, only the st message will be visible as there
number.
ages along with their possible cause
Number Message String Cause Remedy
0 "No error" No errors in queue
-100 "Command error" Unable to complete
requested operation
Unit may be in a mode inconsistent w
request.
ith
-102 "Syntax error"
incorrect.
Command syntax Misspelled or unsupported command
-103 "Invalid separator" SCPI separator not PI section of programming
recognized
See SC
manual.
-104 "Data type error" Data type invalid. Check command for supported data
types
-108 "Parameter not allowed" re additional
ed.
One or mo orrect
parameters were receiv
Check programming manual for c
number of parameters
-109 "Missing parameter" Too few parameters mming manual for correct
received for requested
operation
Check progra
number of parameters
-110 "Command header error" r
rect
Command heade ax of command.
incor
Check synt
-111 "Header separator error" Invalid command
separator used.
Use semi-colon to separate command
headers
-112 "Program mnem
long"
onic too Syntax error Check programming manual for correct
command syntax
-113 "Undefined header" Command not recognized
error
Check programming manual for correct
command syntax
-120 "Numeric data error" Data received is not a
number
Check programming manual for correct
command syntax
-121 "Invalid character in
number"
Number received contains l for correct
non-numeric character(s)
Check programming manua
command syntax
-123 "Exponent too large" Exponent in number
exceeds limits
Check programming manual for correct
parameter range
-128 "Numeric data not
allowed"
hen
number is not allowed.
Check programming manual for correct
command syntax
Number received w
-168 "Block data not allowed" Block data was sent. Check programming manual for correct
command syntax
-200 "Execution error" Command could not be Command may be inconsistent with
California Instruments 146
User Manual Compact i/iX Series
Number Message String Cause Remedy
executed mode of operation.
-201 "Invalid while in local" Command issued but unit Put instrument in remote state before
is not in remote state issuing GPIB commands.
-203 and protected" locked out
gs and are not user accessible.
"Comm Command is Some commands are supported by the
unit but are locked out for protection of
settin
-210 "Trigger error" Problem with trigger
system. measurement.
Unit could not generate trigger for
transient execution or
-211 "Trigger ignored" Trigger request has been
ignored. k
trigger
Trigger setup incorrect or unit was not
armed when trigger was received. Chec
transient system or measurement
system settings.
-213 "Init ignored" Initialization request has ate but
been ignored by
Unit was told to go to armed st
was unable to do so. Could be caused
incorrect transient system or
measurement acquisition setup.
-220 "Parameter error" ed. value.
ramming manual for
Parameter not allow Incorrect parameter or parameter
Check prog
allowable parameters
-221 med
de. transient mode.
As result of *TST? execution, indicates
ALC mode is off or waveform not set to
ine.
"Setting conflict" Transient program
with more than 1 mo
Check other settings. E.g. Redefine
S
-222 f
range. alues
"Data out of range" Parameter data outside o
allowable
Check programming manual for
allowable parameter v
-223 "Too much data" More data received than Check programming manual for number
expected of parameters or data block size
-224 "Illegal parameter value" Parameter value is not anual for correct
中国航空网 www.aero.cn