The same screen also displays the internal temperature of the power supply.
ENTRY DESCRIPTION
ETIME 01:23:45 The ETIME field displays the total
accumulated elapsed time for the instrument
since it's initial manufacture. This value
cannot be changed or reset.
TEMP 37.342° The TEMP field is not a user selectable
parameter but rather a read-out of the internal
temperature in degree
s Celsius. It is provided
for informational purposes only.
California Instruments 76
User Manual Compact i/iX Series
4.2.14 LIMIT Menu
, frequency and current The Limit menu displays the maximum available value for voltage range of the
power supply. This screen is used for information only and contains no user changeable fields. The
limit values shown cannot be changed.
ENTRY DESCRIPTION
LIM LVOLT Low Voltage Displays maximum available output voltage in
Range the low voltage range.
LIM HVOLT High Voltage Displays maximum available output voltage in
Range the high voltage range.
LIM LFREQ Low Frequency
Limit
Displays minimum available output frequency.
LIM HFREQ High Frequency
Limit
Displays maximum available output frequency
CURR C range Displays maximum available current in low
voltage range at full-scale voltage.
P
value indicates split (2) phase configuration.
HASE (C) Phase Setting Displays phase angle for phase C. Valid values
are 120 for three-phase or mode configuration,
0 for single-phase only configuration. Any other
California Instruments 77
User Manual Compact i/iX Series
4.3 Output Programming
4.3.1 Set the Output
Output parameters are all set from the PROGRAM screen.
1. Use the MENU key and select the PROGRAM entry.
2. Press the ENTER key to bring up the PROGRAM menu.
or
2. Use the SET key to directly bring up the PROGRAM menu.
There are two methods for programming output parameters:
IMMEDIATE mode
4.3.2 Slewing Output Values in IMMEDIATE Mode
SET mode
The default mode of operation is an immediate mode in which changes to output parameters made
with the knob or the entry keypad are immediately reflected at the output.
To change the output voltage:
lue
To change the output frequency:
1. Place the cursor on the VOLT entry
2. Rotate the shuttle knob clockwise to increase the value, counterclockwise to decrease the va
or use the Keypad to enter a value and press the Enter key.
These changes take effect immediately.
1. Place the cursor on the FREQ entry
2. Rotate the shuttle knob clockwise to increase the value, counterclockwise to decrease the value
or use the keypad to enter a value and press the Enter key.
These changes take effect immediately.
California Instruments 78
User Manual Compact i/iX Series
4.3.3 Change Output Values in SET Mode
The SET mode of operation is a mode in which changes to output parameters made with the knob or
the entry keypad do not affect the output until the Enter key is pressed. The AC source is put in this
SET mode by pressing the Set key twice. A blinking cursor indicates SET mode is active.
To change the output voltage:
1. Press the Set key twice
2. Place the cursor on the VO
d.
ttle knob clockwise to increase the value, counterclockwise to decrease the value
alues are now present at the output. The unit has returned to
immediate mode of operation until the SET key is pressed again.
as voltage and frequency can be changed from the measurement
LT entry
3. Rotate the shuttle knob clockwise to increase the value, counterclockwise to decrease the value
or enter a new value using the keypad but do not press the Enter key yet.
4. A blinking underline cursor will appear in the data for the VOLT field to indicate a change in
settings but the output remains unchange
5. Place the cursor on the FREQ entry using the down arrow key.
6. Rotate the shu
or enter a new value using the keypad but do not press the Enter key yet.
7. A blinking underline cursor will appear in the data for the FREQ field to indicate a change in
settings but the output remains unchanged.
8. Press the Enter key.
Both new voltage and frequency output v
Note that output settings such
screen as well. If all three phases are selected on three phase models, slewing the shuttle knob will
change the output voltage on all three phases. If only one phase is selected, only the output of the
selected phase will be affected.
California Instruments 79
User Manual Compact i/iX Series
4.4 Waveform Management
The iX Series employs independent arbitrary waveform generators for each phase. This allows the
user to create custom waveforms. In addition, three standard waveforms are always available. This
 