ALC STATE ON GWAddress NC
VOLT REF INT 100.10.1.1
# OUTPUTS THR HostBits 8 NC<
ST PHASE RAND Port No 5025
MEASUREMENT 100.00V 2.349A OUTP CAL VIEW ANGLE -5
TRANSIENT 400.0HZ 234.9W MEAS CAL CAL PWORD 100
238.7VA 1.00pF VOLT FS 2654
2.349A 0.98cF PHASE OFST 0.0
0.18%A 10.559A MVOLT FS 4000
0.19%V 0.0° MCURR FS 5600
TRAN ST IDLE
COUNT 10 APPLICATIONS MIL704
OPTIONS DO160
TRIG SOUR IMM
TRAN STEP AUTO MS704
WHM
California Instruments 56
User Manual Compact i/iX Series
VOLT #12 120.0 LANGUAGE ON
VSLEW #12 1e+06 ADVANCE ON
FREQ #12 400.0 MODE N/A
FSLEW #12 10.00 CLOCK/LOC N/A
DWELL #12 10.00 MIL704 N/A
TTLT #12 ON DO160 N/A
FUNC #12 0 MS704 N/A
PHASE#12 0.0 ABD N/A
CURR #12 11.00 WHM N/A
MB N/A
Power on ETIME/TEMP ETIME 21:20:03
Screens LIMITS TEMP 25:33°C
#1 Initialization
in progress LIM LVOLT 150.0
HVOLT 300.0
#1 CAL. INST.
CIC918-1,Rev0.25 LIM LFREQ 16.00
HFREQ 1000
#2 MODEL 1501iX
SERIAL #54321 CURR 10.00
PHASE(C) 0.0
#3 Memory test
passed
Table 4-1: Menu Tree
California Instruments 57
User Manual Compact i/iX Series
4.2.4 PROGRAM Menu
Figure 4-5: PROGRAM Menu
hown in Figure 4-The PROGRAM menu is s 5. It can be reached in one of two ways:
1. By selecting the MENU r key.
2. By pressing the SET k
The PROGRAM menu is u parameters
are located in the CONTR
The following choices are available in the PROGRAM menus:
key, selecting the PROGRAM entry and pressing the Ente
ey.
sed to change primary output parameters. Less frequently used
OL menu.
ENTRY DESCRIPTION
VOLTAGE
e shuttle (if the voltage field is selected).
Programs the output voltage in Vrms. The voltage can be
changed from 0 to its max range value as determined by the
configuration settings and the selected voltage range using the
keypad + Enter or th
FREQ Programs the output frequency. The frequency can be changed
from its min to its max value as determined by the configuration
settings using the keypad + Enter or the shuttle (if the frequency
field is selected).
VRANGE l
range values may be different depending on the configuration.
The value of this field can be changed with the shuttle as long
as the active pointer () points to the VRANGE entry. If only
one voltage range is available, this field cannot be changed.
Selects 150V or 300V voltage range (if available). The actua
PHASE Selects the phase angle between the external clock and the
output of the AC source. If the clock source is internal, this
parameter has no effect.
FUNC Selects the waveform for the selected phase. On 751iX/1501iX
models, available choices are SINUSOID, SQUARE and
CLIPPED or any user defined waveform that was downloaded to
the AC source waveform memory using the IEEE-488, LAN or
USB interface.
This field is fixed to SINUSOID on 751i/1501i models.
CLIP LEVEL Sets the clip level for the CLIPPED sine wave in percent VTHD.
The range is 0 to 20 %. (751iX/1501iX models only).
Note: Changing the clip level setting will result in temporary loss
of the output voltage as the new clipped waveform is loaded.
This may cause the EUT to reset or turn off. To avoid this, set
the desired clip level before programming the AC voltage and
turning on the output to the EUT or use the transient list system
California Instruments 58
User Manual Compact i/iX Series
ENTRY DESCRIPTION
to switch between waveforms.
VOLT MODE Selects the available output modes of operation. Available
modes are AC, DC (i/iX models) and ACDC (iX models only).
The shuttle can be used to select the desired output mode.
DC OFFSET This parameter applies only when the power source is in ACDC
mode. The DC offset can only be set in percent of the AC RMS
voltage programmed. (Relative programming only). The
available dc offset range is ±20%.
ing the AC voltage and turning on the output to the
EUT.
Note: Changing the offset percentage setting will result in
temporary loss of the output voltage as the new offset is
recalculated and loaded. This may cause the EUT to reset or
turn off. To avoid this, set the desired offset percentage before
programm
CURR Sets the current limit value for the current detection system.
When the load current value exceeds the set current limit, a fault
condition is generated. The actual response of the AC Source to
a current limit fault is determined by the protection mode
selected in the OL MODE field. (CC = Constant Current, CV =
Constant Voltage).